Multimeter

Home > Products > DC/AC Multimeter > LCR Multimeter
TES-2700

LCR Multimeter

Product Features
  • Autoranging and Auto-power-off
  • 3200 count LCD with analog bar-graph
  • Data-hold mode available at ohms.
  • Overload protection up to 600 Vrms
LCR Multimeter

Electrical Specifications :

Accuracies aqre ± (…% of reading + …digits) at 23± 5 and below 80% RH.

DC Voltage : ( Autoranging & Manual range )

Range

Resolution

Accuracy

Input Impedance

Overload Protection

320mV

0.1mV

0.5%+2

> 100MΩ

AC/DC 750V

3.2V

1mV

11MΩ

32V

10mV

10MΩ

320V

100mV

600V

1V

0.8%+2

AC Voltage : 40Hz400Hz ( Autorange & Manual range )

Range

Resolution

Accuracy

Input Impedance

Overload Protection

3.2V

100uV

1.2%+4

11MΩ

AC/DC 750V

32V

1mV

10MΩ

320V

10mV

600V

1V

DC Current : (uA mA Autoranging & Manual Range) 

Range

Resolution

Accuracy

Burden Voltage

Overload Protection

320uA

0.1uA

1%+2

0.2V max

0.5A/250V fast blow fuse & Diode

3200uA

1uA

1.8V max

32mA

10uA

1.2V max

320mA

100uA

1.8V max

20A

10mA

1.5%+2

0.6V max

20A/380V fast blow fuse
10A Continuity
20A for 30 sec. max.

 AC Current : 40Hz400Hz ( uA mA Autorange & Manual range )

Range

Resolution

Accuracy

Burden Voltage

Overload Protection

320uA

0.1uA

1.5%+5

0.2V max

0.5A/250V fast blow fuse & Diode

3200uA

1uA

1.8V max

32mA

10uA

1.2V max

320mA

100uA

1.8V max

20A

10mA

2.0%+5

0.6V max

20A/380V fast blow fuse
10A Continuity
20A for 30 sec. max.

Resistance: ( Autoranging & Manual range )

Range

Resolution

Accuracy

Open Circuit Voltage

Overhead protection

320Ω

0.1Ω

0.8%+ 2

1.3V

660Vrms

3.2KΩ

1Ω

32KΩ

10Ω

320KΩ

100Ω

3.2MΩ

1KΩ

30MΩ

10KΩ

1.5%+ 3


Diode Test: (  )

Range

Resolution

Short Circuit Current

Accuracy

Open Circuit Voltage

Overload Protection

1mV

0.8mA typical

1%+ 2

3V

660Vrms

Audible Continuity:

Range

Resolution

Max Open Circuit Voltage

Overload Protection

20Ω

1.3V

660Vrms


General Specifications:

1.   Display: 3-1/2 digit liquid crystal display (LCD)
maximum reading 3200 and Bargraph indication.

2.   Palarity: Automatic “ -” negative polarity indication.

3.    Overrange:  “OL” appears on the display.

4.   Sampling Rate: 12.5 times/sec of Bar-graph indication,

      2 times /sec of digital display

5.   Power: 2 pcs of  “AAA”  1.5V Battery

6.   Operating Storage Temperature and Humidity :
0 to 40, below 80% RH,
-10 to 60
, below70% RH.

7.   Dimensions : 180(L) x 82(W) x 38(H)mm

8.   Weight : 365g.

9.   Accessories: Test leads (pair), Spare fuse (0.5A/250V),